Vortex
Versatile flowmeter with detection of wet steam conditions and best-in-class accuracy.
Easy energy management – integrated temperature and pressure measurement for steam and gases.
Suitable for a wide range of applications; optimized for steam applications.
Space-saving engineering – inlet run compensation. Same accuracy down to Re 10 000 – most linear Vortex meter body. Long-term stability – robust drift-free capacitive sensor.
Wet steam capabilities for DN 25 to 300 (1 to 12"). Flexible positioning of pressure cell. Industrial siphon design for pressure measurement.
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
Display module with data transfer function. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area).
DN 15 to 300 (½ to 12")
Measuring tube: 1.4408 (CF3M); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4404 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Process connection: 1.4404/F316/F316L); 2.4602
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
Volume flow (liquid): ±0.75 %
Volume flow (optional): ±0.65 %
Volume flow (steam, gas): ±1.00 %
Mass flow (saturated steam): ±1.7% (temperature compensated); ±1.5% (temperature/pressure compensated)
Mass flow (superheated steam, gas): ±1.5 (temperature/pressure compensated); ±1.7% (temperature compensated + external pressure compensation)
Mass flow (liquid): ±0.85%
Liquid: 0.2 to 2100 m³/h (0.15 to 1520 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 1.5 to 28000 m³/h (0.9 to 16600 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)
PN 100, Class 600, 20K
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F)
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F)
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
AlSi10Mg, coated; 1.4404 (316L)
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure
4‐line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available
4‐20 mA HART (passive)
4‐20 mA (passive)
Pulse/frequency/switch output (passive)
Current Input 4‐20 mA (passive)
HART, PROFIBUS PA, FOUNDATION Fieldbus, PROFINET over Ethernet-APL, Modbus TCP over Ethernet-APL
DC 12 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4‐20 mA HART, 4‐20 mA)
DC 12 to 35 V (4‐20 mA HART, pulse/frequency/switch output, 4‐20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)
ATEX, IECEx, cCSAus, JPN, EAC
CE, C-TICK, EAC, UK Ex
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
ABS, LR, BV, DNV, CCS
PED, CRN, AD 2000
3.1 material
NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614‐1, similar to ASME IX (on request)